Cytek Aurora

Cytek Aurora Spectral Flow Cytometer (369 Jones)

5 Lasers: 355, 405, 488, 561, and 640nm

Parameters: 64 fluorescent detectors, plus SSC, vSSC and FSC

The Cytek Aurora is a high-parameter spectral cytometer capable of detecting up to 64 fluorochromes simultaneously. Using its array of 64 detectors, the Aurora examines the entire emission spectra of fluorochromes and uses linear unmixing to investigate the individual parameters. This allows for the combination of highly similar dyes with overlapping emission peaks that would be otherwise indistinguishable on a conventional cytometer. With the 405nm Side Scatter detector (vSSC), this instrument is ideal for the detection of particles as small as 80nm.
View or Download Aurora Configuration

Symphony-A5-Banner

BD Fortessa X-50 (341 Jones)

5 Lasers: 355, 405, 488, 532, and 628nm
Parameters: 28 fluorescent detectors, plus FSC and SSC

The Fortessa X-50 (aka FACSymphony A5) is a five-laser conventional flow cytometer capable of simultaneously detecting up to 28 distinct fluorochromes.
View or Download Fortessa-X50 Configuration

BD FACSymphony A5-SE Spectral with HTS Plate Loader (369 Jones):

5 Lasers: 349, 405, 488, 561, and 637nm

Parameters: 48 fluorescent detectors, plus FSC and SSC

The FACSymphony A5-SE is a high-parameter spectral cytometer that can be used in both spectral and conventional modes, and is capable of detecting 48 fluorescent parameters simultaneously. This offers cross-platform compatibility for the panels designed on other conventional BD cytometers, including our Fortessa and Canto platforms. Like Cytek Aurora spectral analyzer, A5-SE processes entire emission spectra of fluorochromes (in spectral mode) and uses linear unmixing to investigate individual parameters. Therefore, highly similar dyes with overlapping emission peaks can be combined within a panel, that would otherwise be indistinguishable on a conventional cytometer.
View or Download A5-SE Configuration

BD LSRFortessa X-20 (369 Jones)

5 Lasers: 355, 405, 466, 561, and 640nm
Parameters: 18 fluorescent detectors, plus FSC and SSC

The Fortessa X-20 is a five-laser conventional flow cytometer capable of detecting up to 18 distinct fluorochromes simultaneously.
View or Download FortessaX20 Configuration

BD FACSCanto-II (Canto-B with HTS Option; 343 Jones)

3 Lasers: 405, 488, 633nm
Parameters: 8 fluorescent detectors plus FSC and SSC.

FACSCanto II is a three-laser conventional flow cytometer capable of detecting up to 8 fluorescent parameters simultaneosly. Canto B has a highthroughput sampler (HTS) option to allow sample acquisition through regular FACS tubes or directly from a 96 well plate. Additional training is required to operate the HTS loader.
View or Download Canto-A/B Configuration

BD FACSCanto-II (Canto-A; 343 Jones)

3 Lasers: 405, 488, 633nm
Parameters: 8 fluorescent detectors plus FSC and SSC.

Canto-A has similar configuration as the Canto-B except it doesn't have an HTS loader.
View or Download Canto-A/B Configuration

Analyzer Comparison Chart: